Your Cart |
0 items - $0.00 |
Check Out |
Fuel Flow Tester Instructions
Symptom
|
Problem
|
Red LED indicator on fuel flow tester does not light.
|
Check the fuel flow cable for continuity from the instrument to the transducer end.
FS-450: 12 volts between pins 5 and 3 on FF connector.
EDM-700: 12 volts between pins 5 and 6 on FF connector.
EDM-760: 12 volts between pins 5 and 6 on FF connector.
EDM-900/930/1000C: 12 volts between pins 5 and 6 on theP4 FF connector.
|
The FS-450 or EDM displays a positive rate of fuel flow with the fuel flow tester, but not with the fuel flow transducer.
If the K-factor is approximately 19,000 (19.00=KF), the correct rate shown on the display should be between 18 and 20 GPH.
If the K-factor is approximately 29,000 (29.00=KF), the correct rate shown on the display should be between 12 and 14 GPH.
|
If the reading on the display is in the correct rate range, then the fuel flow transducer is most likely defective. Return for repair or replacement. The instrument is okay.
**If some other K-factor transducer is used, the display should read approximately the value resulting from dividing 360 by the K-factor. For example: 360/85.00 = 4.2 displayed.
|
The FS-450 or EDM displays a zero or erroneous rate of fuel flow with the fuel flow tester as well as with the fuel flow transducer.
|
Check that the K-factor is correct. If it is, the instrument is most likely defective. Return the instrument only—not the transducer—to the factory for repair.
|